62 •
Channels Within each wireless group, the devices communicate
with each other on three channels (C1, C2, C3).
When you set up a wireless group with this device, it
suggests three channels guaranteeing optimum data
transmission. We recommend accepting the channel
numbers suggested.
You can also select the channel numbers (0 - 99)
manually - for example if you want to set up more than
one wireless group.
The channels must be sufficiently far apart to ensure
trouble-free data transmission. We recommend a
spacing of at least 30. Each channel number may only
be used for one channel.
Example of configuration; channel numbers when
setting up 3 wireless groups within one surgery:
wireless group 0: C1=_0, C2= 30, C3=60
wireless group 1: C1=10, C2=40, C3=70
wireless group 2: C1=20, C2=50, C3=60
Detection of devices If you wish to set up a wireless group with this device, it
searches for other active devices from the
seca 360° wireless system. The devices detected are
shown by numbers in the form of modules in the device
display (e.g. MO 3). The meaning of the numbers is as
follows:
1: personal scales
2: height-measuring rod
3: wireless printer
4: PC with seca USB wireless module
7: baby scales
5, 6 and 8-12: reserved for system expansion